Noise and Health (Jan 2017)
Effect of Filters on the Noise Generated by Continuous Positive Airway Pressure Delivered via a Helmet
Abstract
Introduction: One of the problems that the delivery of continuous positive airway pressure (CPAP) via a helmet poses is the generation of noise. The objective of our study was to assess the effect that the use of filter has on sound pressure levels generated by the delivery of positive airway pressure at different gas flow rates. Materials and Methods: Sound pressure levels generated by neonatal helmet CPAP delivery were measured at different gas flows (20, 30, and 40 l/min) with and without a breathing filter. Noise intensity was measured by installing microphones in the inner ear of dummy heads wearing helmets. Results: The sound pressure level increased by 38% at a gas flow of 40 l/min, as compared to a gas flow of 20 l/min {74 dBA [interquartile range (IQR) 2,2] vs 52 dBA (IQR 5,9), respectively}. Using the breathing filter as a diffuser has a variety of effects on sound pressure levels according to the gas flow rate. Conclusion: The intensity of the noise generated by helmet delivery of positive airway pressure depends on the type of helmet used, gas flow, and use or not of a diffuser filter. Breathing filters with gas flows over 30 l/min might not be recommended since they would not attenuate but will rather amplify sound pressure.
Keywords