Scientific Reports (Jun 2022)

Direct in-water radiation dose measurements using Cherenkov emission corrected signals from polarization imaging for a clinical radiotherapy application

  • Émily Cloutier,
  • Luc Beaulieu,
  • Louis Archambault

DOI
https://doi.org/10.1038/s41598-022-12672-w
Journal volume & issue
Vol. 12, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Cherenkov emission (CE) is a visible blueish light emitted in water mediums irradiated by most radiotherapy treatment beams. However, CE is produced anisotropically which currently imposes a geometrical constraint uncertainty for dose measurements. In this work, polarization imaging is proposed and described as a method enabling precise 2D dose measurements using CE. CE produced in a water tank is imaged from four polarization angles using a camera coupled to a rotating polarizer. Using Malus’ law, the polarized component of CE is isolated and corrected with Monte Carlo calculated CE polar and azimuthal angular distributions. Projected dose measurements resulting from polarization-corrected CE are compared to equivalent radiochromic film measurements. Overall, agreement between polarized corrected CE signal and films measurements is found to be within 3%, for projected percent depth dose (PPDD) and profiles at the different tested energies ( $$\gamma$$ γ : 6 and $$18\,\hbox {MV}$$ 18 MV , e $$^{-}$$ - : 6 and 18 $$\,\hbox {MeV}$$ MeV ). In comparison, raw Cherenkov emission presented deviations up 60% for electron beam PPDDs and 20% for photon beams PPDDs. Finally, a degree of linear polarization between 29% and 47% was measured for CE in comparison to $$0.2\pm 0.3$$ 0.2 ± 0.3 % for scintillation. Hence, polarization imaging is found to be a promising and powerful method for improved radio-luminescent dose measurements with possible extensions to signal separation.