Measurement: Sensors (Aug 2022)

The 20 N⋅m dynamic torque calibration device at PTB

  • Leonard Klaus,
  • Michael Kobusch

Journal volume & issue
Vol. 22
p. 100370

Abstract

Read online

Research in the field of dynamic torque calibration has gained some momentum in the last few years. At PTB, a calibration device for small dynamic torque levels up to about 0.5 N⋅m proved the working principle of dynamic torque calibrations with sinusoidal excitation. In a next step, a new calibration device was developed with the intention of generating higher dynamic torque levels in the range of up to 20 N⋅m. Although this is still less than many industrial applications require, it could pave the way to higher torque levels. The new calibration device features a horizontal arrangement of all components of the drive shaft. The device consists mostly of industrial components that have been modified, rather than custom-made components. After commissioning, optimising the device enabled a frequency range from 10 Hz to up to 400 Hz with dynamic torque amplitudes of up to 20 N⋅m.

Keywords