EJNMMI Physics (Jun 2019)

Augmenting camera images with gamma detector data

  • Peter A. von Niederhäusern,
  • Simon Pezold,
  • Uri Nahum,
  • Carlo Seppi,
  • Guillaume Nicolas,
  • Michael Rissi,
  • Stephan K. Haerle,
  • Philippe C. Cattin

DOI
https://doi.org/10.1186/s40658-019-0245-z
Journal volume & issue
Vol. 6, no. 1
pp. 1 – 21

Abstract

Read online

Abstract Background Squamous cell carcinoma in the head and neck region is one of the most widespread cancers with high morbidity. Classic treatment comprises the complete removal of the lymphatics together with the cancerous tissue. Recent studies have shown that such interventions are only required in 30% of the patients. Sentinel lymph node biopsy is an alternative method to stage the malignancy in a less invasive manner and to avoid overtreatment. In this paper, we present a novel approach that enables a future augmented reality device which improves the biopsy procedure by visual means. Methods We propose a co-calibration scheme for axis-aligned miniature cameras with pinholes of a gamma ray collimating and sensing device and show results gained by experiments, based on a calibration target visible for both modalities. Results Visual inspection and quantitative evaluation of the augmentation of optical camera images with gamma information are congruent with known gamma source landmarks. Conclusions Combining a multi-pinhole collimator with axis-aligned miniature cameras to augment optical images using gamma detector data is promising. As such, our approach might be applicable for breast cancer and melanoma staging as well, which are also based on sentinel lymph node biopsy.

Keywords