BMC Pediatrics (May 2020)

Diagnostic test accuracy of new generation tympanic thermometry in children under different cutoffs: a systematic review and meta-analysis

  • Dan Shi,
  • Li-Yuan Zhang,
  • Hai-Xia Li

DOI
https://doi.org/10.1186/s12887-020-02097-7
Journal volume & issue
Vol. 20, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Background The infrared tympanic thermometer (IRTT) is a popular method for temperature screening in children, but it has been debated for the low accuracy and reproducibility compared with other measurements. This study was aimed to identify and quantify studies reporting the diagnostic accuracy of the new generation IRTT in children and to compare the sensitivity and specificity of IRTT under different cutoffs and give the optimal cutoff. Methods Articles were derived from a systematic search in PubMed, Web of Science Core Collection, and Embase, and were assessed for internal validity by the Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2). The figure of risk of bias was created by Review Manager 5.3 and data were synthesized by MetaDisc 1.4. Results Twelve diagnostic studies, involving 4639 pediatric patients, were included. The cut-offs varied from 37.0 °C to 38.0 °C among these studies. The cut-off 37.8 °C was with the highest sROC AUC (0.97) and Youden Index (0.83) and was deemed to be the optimal cutoff. Conclusion The optimal cutoff for infrared tympanic thermometers is 37.8 °C. New Generation Tympanic Thermometry is with high diagnostic accuracy in pediatric patients and can be an alternative for fever screening in children.

Keywords