Micromachines (Dec 2014)

Coplanar Electrode Layout Optimized for Increased Sensitivity for Electrical Impedance Spectroscopy

  • Casper Hyttel Clausen,
  • Gustav Erik Skands,
  • Christian Vinther Bertelsen,
  • Winnie Edith Svendsen

DOI
https://doi.org/10.3390/mi6010110
Journal volume & issue
Vol. 6, no. 1
pp. 110 – 120

Abstract

Read online

This work describes an improvement in the layout of coplanar electrodes for electrical impedance spectroscopy. We have developed, fabricated, and tested an improved electrode layout, which improves the sensitivity of an impedance flow cytometry chip. The improved chip was experimentally tested and compared to a chip with a conventional electrode layout. The improved chip was able to discriminate 0.5 μm beads from 1 μm as opposed to the conventional chip. Furthermore, finite element modeling was used to simulate the improvements in electrical field density and uniformity between the electrodes of the new electrode layout. Good agreement was observed between the model and the obtained experimental results.

Keywords