Practical Laboratory Medicine (Mar 2024)

Assessment of novel POCT to evaluate liver function

  • Carmen Minea,
  • Damien Gruson

Journal volume & issue
Vol. 39
p. e00367

Abstract

Read online

Objectives: Point of care testing (POCT) offers the possibility of near bedside patient testing with a reduction of the turn-around time of analysis. The aim of our study was to determine the analytical performances and usability of a recently developed POCT device for the measurement of tests related to liver function. We evaluated the performance of a liver tests panel performed on the LINX EVO® POCT device. Design and methods: The imprecision was determined with the Bio-Rad Liquichek Unassayed Chemistry Control. Method comparison was performed with a Cobas® 8000 analyzer. Samples from twenty healthy volunteers were used to verify the reference intervals. Furthermore, practicality was assessed by the healthcare staff handling the POCT device through a dedicated questionnaire. Results: The imprecision observed was matching the criteria for the in-lab assay with only one exception, globulin, with an observed imprecision of 6.3 % and a criteria of 5.7 %. With the exception of total and direct bilirubin, the POCT method showed good agreement with the in-lab methods. The verification of reference intervals showed that more than 90 % of the healthy volunteer values were included into the reference interval claimed by the manufacturer except for glucose and globulin. The POCT practicality questionnaire was satisfying overall for users. Conclusions: Our study showed very good analytical performances overall for the liver test panel performed on the LINX EVO® POCT instrument.