Sensors (Apr 2024)

XMEA: A New Hybrid Diamond Multielectrode Array for the In Situ Assessment of the Radiation Dose Enhancement by Nanoparticles

  • Patricia Nicolucci,
  • Guilherme Gambaro,
  • Kyssylla Monnyelle Araujo Silva,
  • Iara Souza Lima,
  • Oswaldo Baffa,
  • Alberto Pasquarelli

DOI
https://doi.org/10.3390/s24082409
Journal volume & issue
Vol. 24, no. 8
p. 2409

Abstract

Read online

This work presents a novel multielectrode array (MEA) to quantitatively assess the dose enhancement factor (DEF) produced in a medium by embedded nanoparticles. The MEA has 16 nanocrystalline diamond electrodes (in a cell-culture well), and a single-crystal diamond divided into four quadrants for X-ray dosimetry. DEF was assessed in water solutions with up to a 1000 µg/mL concentration of silver, platinum, and gold nanoparticles. The X-ray detectors showed a linear response to radiation dose (r2 ≥ 0.9999). Overall, platinum and gold nanoparticles produced a dose enhancement in the medium (maximum of 1.9 and 3.1, respectively), while silver nanoparticles produced a shielding effect (maximum of 37%), lowering the dose in the medium. This work shows that the novel MEA can be a useful tool in the quantitative assessment of radiation dose enhancement due to nanoparticles. Together with its suitability for cells’ exocytosis studies, it proves to be a highly versatile device for several applications.

Keywords