Sensors (Nov 2024)

Noninvasive Early Detection of Systemic Inflammatory Response Syndrome of COVID-19 Inpatients Using a Piezoelectric Respiratory Rates Sensor

  • Tsuyoshi Kobayashi,
  • Takemi Matsui,
  • Isamu Sugita,
  • Norihiro Tateda,
  • Shohei Sato,
  • Kenichi Hashimoto,
  • Masei Suda

DOI
https://doi.org/10.3390/s24227100
Journal volume & issue
Vol. 24, no. 22
p. 7100

Abstract

Read online

In 2020, 20% of patients with COVID-19 developed severe complications, including life-threatening pneumonia with systemic inflammatory response syndrome (SIRS). We developed a preliminary SIRS monitor that does not require blood sampling, is noninvasive, and can collect data 24 h per day. The proposed monitor comprises a piezoelectric respiratory sensor located beneath the patient’s mattress and a fingertip pulse sensor that determines ultra-high accuracy respiratory rate (mode of a 40-min frequency distribution of respiratory rates (M40FD-RR)). We assessed the clinical performance of the M40FD-RR preliminary SIRS monitor in 29 patients (12 female, 17 male, aged 15–90 years) hospitalized at Suwa Central Hospital with COVID-19, which was confirmed by a positive polymerase chain reaction test. SIRS was evaluated by logistic regression analysis using M40FD-RR, heart rate, age, and sex as explanatory variables. We compared the results of 109 examinations of 29 COVID-19 inpatients with SIRS against those determined by the proposed monitor. The proposed monitor achieved 75% sensitivity and 83% negative predictive value, making it a promising candidate for future 24 h noninvasive preliminary SIRS tests.