Pharmaceutics (Mar 2020)

Optimal Connection for Tiotropium SMI Delivery through Mechanical Ventilation: An In Vitro Study

  • Tien-Pei Fang,
  • Yu-Ju Chen,
  • Tsung-Ming Yang,
  • Szu-Hu Wang,
  • Ming-Szu Hung,
  • Shu-Hua Chiu,
  • Hsin-Hsien Li,
  • James B. Fink,
  • Hui-Ling Lin

DOI
https://doi.org/10.3390/pharmaceutics12030291
Journal volume & issue
Vol. 12, no. 3
p. 291

Abstract

Read online

We aimed to quantify Soft Mist Inhalers (SMI) delivery to spontaneous breathing model and compare with different adapters via endotracheal tube during mechanical ventilation or by manual resuscitation. A tiotropium SMI was used with a commercial in-line adapter and a T-adapter placed between the Y-adapter and the inspiratory limb of the ventilator circuit during mechanical ventilation. The SMI was actuated at the beginning of inspiration and expiration. In separate experiments, a manual resuscitator with T-adapter was attached to endotracheal tube, collecting filter, and a passive test lung. Drug was eluted from collecting filters with salt-based solvent and analyzed using high-performance liquid chromatography. Results showed the percent of SMI label dose inhaled was 3-fold higher with the commercial in-line adapter with actuation during expiration than when synchronized with inspiration. SMI with T-adapter delivery via ventilator was similar to inhalation (1.20%) or exhalation (1.02%), and both had lower delivery dose than with manual resuscitator (2.80%; p = 0.01). The inhaled dose via endotracheal tube was much lower than inhaled dose with spontaneous breathing (22.08%). In conclusion, the inhaled dose with the commercial adapter was higher with SMI actuated during expiration, but still far less than reported spontaneous inhaled dose.

Keywords