Nuclear Technology and Radiation Protection (Jan 2012)

Prototype of the particle fluence measurement device

  • Trajić Ivan M.,
  • Beličev Petar D.,
  • Petrović Srđan M.,
  • Rajčević Milan N.,
  • Munitlak Dragan M.,
  • Bojović Boško P.

DOI
https://doi.org/10.2298/NTRP1204404T
Journal volume & issue
Vol. 27, no. 4
pp. 404 – 408

Abstract

Read online

A method of measuring particle fluence and ion beam current in experiments of ion implantation is presented. The design and technical details of a realized prototype device are discussed. Special care is devoted to the effect of secondary electrons and its elimination. An electrostatic solution with two electrodes for the suppression of secondary electrons has been realized. Our approach provides redundancy and allows lower bias voltages to be used for the suppression of electrons. The presented results illustrate the efficiency of the proposed solution. [Projekat Ministarstva nauke Republike Srbije, br. III 45006]

Keywords