Endoscopy International Open (Sep 2015)

Development of a device for detecting target specimens from EUS-guided FNA samples

  • Kazuya Matsumoto,
  • Masaru Ueki,
  • Yohei Takeda,
  • Kenichi Harada,
  • Takumi Onoyama,
  • Soichiro Kawata,
  • Yuichiro Ikebuchi,
  • Ryu Imamoto,
  • Yasushi Horie,
  • Yoshikazu Murawaki

DOI
https://doi.org/10.1055/s-0034-1393076
Journal volume & issue
Vol. 03, no. 06
pp. E662 – E664

Abstract

Read online

Background and study aims: Specimens collected by fine needle are microscopic and contain blood; therefore, the presence of a target specimen within a sample is often difficult to confirm. Although rapid on-site evaluation (ROSE) during endoscopic ultrasound-guided fine needle aspiration biopsy (EUS-FNA) is beneficial, many health care facilities are unable to apply this technique due to a lack of cytopathologists. The aim of this study was to develop and validate a device that detects the target specimen within pancreatic tumor EUS-FNA samples. Patients and methods: Fifty-eight consecutive patients with solid pancreatic masses were studied for a preliminary case series at a tertiary-care university hospital (Tottori University Hospital, Yonago, Japan). The material collected was checked with a target sample check illuminator (TSCI) and was evaluated by one cytopathologist. Results: The agreement rate between the TSCI and histopathology was 93.7 %. Further testing procedures were not needed in 91.4 % of patients, and the mean number of needle punctures was 1.2 after a single pass using TSCI. No adverse events were encountered with the procedure. Conclusions: With the introduction of the TSCI in EUS-FNA, it became possible to both collect the minimum necessary target samples by EUS-FNA and to end further procedures, even without performing ROSE.