Abstract
Abstract Background Kilohertz frequency alternating current (KHFAC) waveforms reversibly block conduction in mammalian peripheral nerves. The initiation of the KHFAC produces nerve activation, called the onset response, before complete block occurs. An amplitude ramp, starting from zero amplitude, is ineffective in eliminating this onset activity. We postulated that initiating the ramp from a non-zero amplitude would produce a different effect on the onset. Methods Experiments were conducted in an in vivo rat model. KHFAC was applied at supra block threshold amplitudes and then reduced to a lower sub block amplitude (25, 50, 75 and 90% of the block threshold amplitude). The amplitude was then increased again to the original supra block threshold amplitude with an amplitude ramp. This ramp time was varied for each of the amplitude levels tested. Results The amplitude ramp was successful in eliminating a second onset. This was always possible for the ramps up from 75 and 90% block threshold amplitude, usually from 50% but never from 25% of the block threshold amplitude. Conclusions This maneuver can potentially be used to initiate complete nerve block, transition to partial block and then resume complete block without producing further onset responses.