Calibration of Calorimetric Measurement in a Liquid Argon Time Projection Chamber

  • Tingjun Yang

DOI
https://doi.org/10.3390/instruments5010002
Journal volume & issue
Vol. 5, no. 1
p. 2

Abstract

Read online

The liquid argon time projection chamber provides high-resolution event images and excellent calorimetric resolution for studying neutrino physics and searching for beyond-standard-model physics. In this article, we review the main physics processes that affect detector response, including the electronics and field responses, space charge effects, electron attachment to impurities, diffusion, and recombination. We describe methods to measure those effects, which are used to calibrate the detector response and convert the measured raw analog-to-digital converter (ADC) counts into the original energy deposition.