Journal of Vascular Surgery Cases and Innovative Techniques (Jun 2023)

Fiber Optic RealShape imaging using upper extremity and transfemoral access for fenestrated-branched endovascular aortic aneurysm repair

  • Felipe L. Pavarino, MD,
  • Jesus Porras-Colon, MD,
  • Marilisa Soto-Gonzalez, MDM,
  • Alejandro Pizano, MD,
  • Mirza S. Baig, MD,
  • Carlos H. Timaran, MD

Journal volume & issue
Vol. 9, no. 2
p. 101191

Abstract

Read online

We report our initial experience using Fiber Optic RealShape (FORS), an innovative real-time three-dimensional visualization technology that uses light instead of radiation, to achieve upper extremity (UE) access during fenestrated/branched endovascular aortic aneurysm repair (FBEVAR). An 89-year-old male patient with a type III thoracoabdominal aortic aneurysm, unfit for open aortic repair, underwent FBEVAR. Dual fluoroscopy, intravascular ultrasound, and three-dimensional fusion overlay were used, in addition to FORS. All target artery catheterizations were successfully accomplished using FORS, from UE access, without radiation. Our experience demonstrates that FBEVAR with FORS using UE access can be used for target artery catheterization without radiation.

Keywords