Virology Journal (Jul 2019)

CIN2+ detection of the HPV DNA Array genotyping assay in comparison with the Cobas 4800 HPV test and cytology

  • Aleksandra Pesic,
  • Amrei Krings,
  • Matthias Hempel,
  • Rosemarie Preyer,
  • Kimon Chatzistamatiou,
  • Theodoros Agorastos,
  • Andreas M. Kaufmann

DOI
https://doi.org/10.1186/s12985-019-1197-6
Journal volume & issue
Vol. 16, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Background HPV DNA Array is an E1-targeting PCR genotyping test, with capability of distinguishing 18 high-risk (16, 18, 26, 31, 33, 35, 39, 45, 51, 52, 53, 56, 58, 59, 66, 68, 73, 82) and 11 low-risk HPV types (6, 11, 40, 42, 44, 54, 67, 69, 70, 85, 97). HPV DNA Array uses multiplex PCR for E1-gene sequence amplification. The amplicons are detected and genotyped by reverse hybridization to immobilized DNA probes spotted as triplets in single 96 well-plate wells and read by AID ELISPOT reader. Methods Aim of the study was to evaluate the clinical performance of the assay against internationally accepted and FDA approved Cobas 4800 HPV test (Roche Diagnostics). Study population comprised of 500 cervical samples. Results HPV DNA Array demonstrated a very high sensitivity of 100% for CIN2+ and 100% for CIN3+ detection, same as Cobas 4800. HPV DNA Array showed greater sensitivity for CIN2+ detection than cytology (100% vs. 13.6%). The agreement to Cobas 4800 for HPV detection, irrespective of type, was 81.4% with κ = 0.613. The agreement for HPV 16 was 92.8% (κ = 0.929), and for HPV 18 54.2% (κ = 0.681). Conclusion HPV DNA Array demonstrated good clinical performance for detection of high-grade lesions, and may be considered for usage in a screening setting.