Diagnostics (Sep 2022)

Improved Single Breath-Hold SSFSE Sequence for Liver MRI Based on Compressed Sensing: Evaluation of Image Quality Compared with Conventional T2-Weighted Sequences

  • Hyun Kyung Lee,
  • Ji Soo Song,
  • Weon Jang,
  • Dominik Nickel,
  • Mun Young Paek

DOI
https://doi.org/10.3390/diagnostics12092164
Journal volume & issue
Vol. 12, no. 9
p. 2164

Abstract

Read online

The purpose of this study was to evaluate the image quality of compressed-sensing accelerated single-shot fast spin-echo (SSFSECS) sequences acquired within a single breath-hold in comparison with conventional SSFSE (SSFSECONV) and multishot TSE (mTSE). A total of 101 patients who underwent liver MRI at 3 T, including SSFSECONV (acquisition time (TA) = 58–62 s), mTSE (TA = 108 s), and SSFSECS (TA = 18 s), were included in this retrospective study. Two radiologists assessed the three sequences with respect to artifacts, organ sharpness, small structure visibility, overall image quality, and conspicuity of main lesions of liver and pancreas using a five-point evaluation scale system. Descriptive statistics and the Wilcoxon signed-rank test were used for statistical analysis. SSFSECS was significantly better than SSFSECONV and mTSE for artifacts, small structure visibility, overall image quality, and conspicuity of main lesions (p CS did not significantly differ (p = 0.554). Conspicuity of liver lesion did not significantly differ between SSFSECONV and mTSE (p = 0.404). SSFSECS showed superior image quality compared with SSFSECONV and mTSE despite a more than three-fold reduction in TA, suggesting a remarkable potential for saving time in liver imaging.

Keywords