Orthopaedic Surgery (Feb 2019)

Integral Fixation Titanium/Polyetheretherketone Cages for Cervical Arthrodesis: Evolution of Cage Design and Early Radiological Outcomes and Fusion Rates

  • Kevin Phan,
  • Matthew H Pelletier,
  • Prashanth J Rao,
  • Wen Jie Choy,
  • William R Walsh,
  • Ralph J Mobbs

DOI
https://doi.org/10.1111/os.12413
Journal volume & issue
Vol. 11, no. 1
pp. 52 – 59

Abstract

Read online

Objective To evaluate the initial outcomes of a composite cage with integral fixation using the Redmond titanium (Ti)/polyetheretherketone (PEEK) anterior cervical discectomy and fusion (ACDF) device. Methods Data from 50 consecutive patients were prospectively collected from a single senior surgeon cohort. All cages were between 5 and 8 mm in height, and were packed with supercritical CO2 sterilized allograft. Patients were followed up for a minimum of 6 months, and implant complications were assessed. Results From the original cohort, three were unavailable for follow‐up. Forty‐seven patients with a total of 58 operative levels were observed for a mean of 7.9 months. A fusion rate of 96% was achieved. Good to excellent outcomes were seen in 92% of patients. There were no cases of implant Ti/PEEK delamination or implant failure, with excellent early fusion rates using supercritical CO2 allograft. Conclusions The present study demonstrates the development of a composite ACDF cage design that is a safe and effective treatment option with the potential for early osseointegration and interbody fusion. Supercritical CO2 sterilized allograft was an effective graft material supporting fusion.

Keywords