BMC Oral Health (Sep 2023)

Nickel ion release and surface analyses on instrument fragments fractured beyond the apex: a laboratory investigation

  • Sıdıka Mine Toker,
  • Ekim Onur Orhan,
  • Arzu Beklen

DOI
https://doi.org/10.1186/s12903-023-03434-9
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 15

Abstract

Read online

Abstract Background To analyse the changes in surface and nickel ion release characteristics of fractured root canal shaping instruments in a simulated body fluid environment. Methods A total of 54 new instruments were studied. The instrument groups consisted of five different NiTi alloys and a stainless-steel alloy. To standardize instrument fracture, a torsional type of failure was created on each instrument. The fractured specimens of each instrument group were randomly divided into three static immersion subgroups of 1 h, 7-day, and 30-day (n = 3). Simulated body fluid (SBF) was prepared to mimic human blood plasma by Kokubo&Takadama protocol for ex situ static immersions at 37ºC. The surfaces were examined via scanning electron microscopy coupled with energy-dispersive X-ray spectroscopy. To determine the quantitative ion release, the retrieved SBFs were analyzed using inductively coupled plasma mass spectrometry. Two-way ANOVA and Tukey post hoc tests sought the statistical significance of the nickel ion values(p 0.05). Conclusions Within the limitations of the presented study, it was concluded that the fractured SS and NiTi root canal instruments release Ni ions in contact with body fluid. However, the Ni ion release values determined during the observation periods are lower than the critical toxic or allergic thresholds defined for the human body. This was due to the ionic dissolution cycle reaching a stable state from 1-hour to 30-day exposure to the body fluid of fractured instruments.

Keywords