Brazilian Journal of Radiation Sciences (Oct 2019)

Sum-peak method with two NaI(Tl) crystals: 68(Ge+Ga) standardization

  • Estela Maria Oliveira,
  • Akira Iwahara,
  • Roberto Poledna,
  • Carlos José da Silva,
  • Paulo Alberto Lima da Cruz,
  • Regio dos Santos Gomes,
  • José Ubiratan Delgado,
  • Ricardo Tadeu Lopes

DOI
https://doi.org/10.15392/bjrs.v7i3b.862
Journal volume & issue
Vol. 7, no. 3B (Suppl.)

Abstract

Read online

A 68(Ge+Ga) commercial solution has been standardized in LNMRI in Brazil, by sum-peak method, in which a 3”x3” NaI(Tl) gamma-ray detector is positioned at the top of a well-type 5”x5” NaI(Tl) gamma-ray detector, resulting in a set up approximately 4p geometry. In this work the known germanium volatility was tested using three dried sources and three liquid sources in the sum-peak method measurements and the activity results showed a standard deviation of 0.41%. The activities were compared with another primary method: 4pb-g live-timed anti-coincidence counting. The two methods gave activity concentration values with differences from the certified value of +0.8 % (anticoincidence method) and -3.4% (sum-peak method).