Advances in Electrical and Electronic Engineering (Jan 2012)

Ultrasonic Transducer Peak-to-Peak Optical Measurement

  • Pavel Skarvada,
  • Pavel Tofel,
  • Pavel Tomanek

DOI
https://doi.org/10.15598/aeee.v10i2.618
Journal volume & issue
Vol. 10, no. 2
pp. 125 – 129

Abstract

Read online

Possible optical setups for measurement of the peak-to-peak value of an ultrasonic transducer are described in this work. The Michelson interferometer with the calibrated nanopositioner in reference path and laser Doppler vibrometer were used for the basic measurement of vibration displacement. Langevin type of ultrasonic transducer is used for the purposes of Electro-Ultrasonic Nonlinear Spectroscopy (EUNS). Parameters of produced mechanical vibration have to been well known for EUNS. Moreover, a monitoring of mechanical vibration frequency shift with a mass load and sample-transducer coupling is important for EUNS measurement.