Journal of Medical Physics (Jan 2009)

The role of Cobalt-60 in modern radiation therapy: Dose delivery and image guidance

  • Schreiner L,
  • Joshi Chandra,
  • Darko Johnson,
  • Kerr Andrew,
  • Salomons Greg,
  • Dhanesar Sandeep

Journal volume & issue
Vol. 34, no. 3
pp. 133 – 136

Abstract

Read online

The advances in modern radiation therapy with techniques such as intensity-modulated radiation therapy and image-guid-ed radiation therapy (IMRT and IGRT) have been limited almost exclusively to linear accel-erators. Investigations of modern Cobalt-60 (Co-60) radiation delivery in the context of IMRT and IGRT have been very sparse, and have been limited mainly to computer-modeling and treatment-planning exercises. In this paper, we report on the results of experiments using a tomotherapy benchtop apparatus attached to a conventional Co-60 unit. We show that conformal dose delivery is possible and also that Co-60 can be used as the radiation source in megavoltage computed tomography imaging. These results complement our modeling studies of Co-60 tomotherapy and provide a strong motivation for continuing development of modern Cobalt-60 treatment devices.

Keywords