Proceedings (Jan 2019)

Solvent Resistant O2 Sensor Integrated in Pressured Flow Reactors

  • Philipp Sulzer,
  • René Lebl,
  • Torsten Mayr

DOI
https://doi.org/10.3390/proceedings2130994
Journal volume & issue
Vol. 2, no. 13
p. 994

Abstract

Read online

In this contribution we present an optical oxygen sensor that can easily be integrated into continuous flow reactors. The sensors are pressure resistant to up to 8 bars and are able to continuously measure oxygen for several hours in organic solvents. The dynamic range was evaluated up to 42 mmol/L O2 and several calibrations showed the sensors to perform linearly according to Stern-Volmer, which allows convenient two-point calibrations. By using state-of-the-art HPLC connectors and materials the sensor can be implemented at virtually every point of a standard flow reactor. Readout is conducted via a 4-channel USB phase fluorimeter.

Keywords