Frontiers in Cardiovascular Medicine (Feb 2023)

Pulse-wave velocity assessments derived from a simple photoplethysmography device: Agreement with a referent device

  • Gabriel Zieff,
  • Keeron Stone,
  • Craig Paterson,
  • Craig Paterson,
  • Simon Fryer,
  • Jake Diana,
  • Jade Blackwell,
  • Jade Blackwell,
  • Michelle L. Meyer,
  • Lee Stoner

DOI
https://doi.org/10.3389/fcvm.2023.1108219
Journal volume & issue
Vol. 10

Abstract

Read online

ObjectivePulse-wave velocity (PWV), a common measure of arterial stiffness, can be measured continuously and across multiple body sites using photoplethysmography (PPG). The objective was to determine whether a simple photoplethysmography PPG PWV method agrees with a referent device.ApproachPhotoplethysmography heart-finger PWV (hfPWV) and heart-toe PWV (htPWV) were compared to oscillometric carotid-wrist PWV (cwPWV) and carotid-ankle PWV (caPWV) referent measurements, respectively. In 30 adults (24.6 ± 4.8 years, body mass index 25.2 ± 5.9 kg/m2, 18 female), three measurements were made: two supine baseline measurements (Base 1, Base 2) and one measurement (Tilt) 5 min after a modified head-up tilt test (mHUTT). Overall agreement and repeated measures agreement (change in PPG PWV from Base to Tilt vs. change in referent PWV from Base to Tilt) were calculated using linear mixed models. Agreement estimates were expressed as intra-class correlation coefficients (ICC).Main resultsFor hfPWV there was strong overall agreement (ICC: 0.77, 95%CI: 0.67–0.85), but negligible and non-significant repeated measures agreement (ICC: 0.10, 95%CI: −0.18 to 0.36). For htPWV, there was moderate overall agreement (ICC:0.50, 95%CI: 0.31–0.65) and strong repeated measures agreement (ICC: 0.81, 95%CI: 0.69–0.89).SignificancePhotoplethysmography can continuously measure PWV at multiple arterial segments with moderate-strong overall agreement. While further work with upper-limb PPG PWV is needed, PPG can adequately capture acute changes in lower-limb PWV.

Keywords