Breast (Dec 2022)
Application of thermoplastic elastomer (TPE) bolus in postmastectomy radiotherapy
Abstract
Purpose: To assess the planned dose, in vivo dosimetry, acute skin toxicity, pain, and distress using Thermoplastic Elastomer (TPE) bolus for postmastectomy radiotherapy (PMRT). Material and methods: Thirty-two PMRT patients with TPE bolus (17 patients for 25 fractions, 15 patients for the first 20 fractions) were selected for the study. The acute skin toxicity, pain, and psychological distress were assessed from the first treatment week to the fourth week after the end of treatment. At the first treatment, the MOSFET was used in vivo dosimetry measurement. Results: In vivo dosimetry with the bolus, the dose deviation ranged from −6.22% to −1.56% for 5 points. The presence of grade 1 and 2 skin toxicity reached its peak (70.0% and 13.3%) in the sixth week. Two patients (6.6%) with 25 fractions bolus experienced moist desquamation in the fifth and seventh week, with pain score 2 and 3, and interruptions of 3 and 5 days, respectively. The incidence of pain score 1, 2, and 3 peaked in the fifth (33.3%), fourth (33.3%), and seventh (10.0%) week. No patients experienced grade 3 skin toxicity and severe pain. One patient had significant anxiety, and two patients had significant depression. Conclusion: The TPE bolus can accurately fit skin and improve the surface dose to more than 90%. Twenty fractions with TPE bolus had similar skin toxicity and pain to those without bolus and did not increase patients' distress and clinical workload, compared with the literature's data, which is an alternative to the 3D printing bolus for PMRT.