Practical Laboratory Medicine (Aug 2019)

Improved quality of samples and laboratory turnaround time using 3.5 mL low vacuum BD Vacutainer® Barricor tubes in the emergency department

  • S. Badiou,
  • O. Vuillot,
  • A.S. Bargnoux,
  • N. Kuster,
  • S. Lefebvre,
  • M. Sebbane,
  • J.P. Cristol,
  • A.M. Dupuy

Journal volume & issue
Vol. 16

Abstract

Read online

Background: Centrifugation is a consuming time step which participates to increase the turnaround time (TAT) in laboratories, likewise hemolysis sample that needs a re-sampling could delay management of patients. Recently, it has been postulated that BD Barricor™ tube could allow to decrease the centrifugation time and prevent hemolysis, two key feature to ensure high-quality results.Aim of the study was to evaluate the impact of replacing 4 mL BD vacutainer heparin lithium tube by low vacuum 3.5 mL BD vacutainer Barricor™ tube in an emergency department (ED) on hemolysis rate and TAT. Methods: Data of hemolysis index (HI) and TAT were compared between the first period of 15 days using 4 mL BD vacutainer heparin lithium tubes with 15 min at 2000xg as centrifugation setting and a second period of 15 days using BD vacutainer Barricor™ tube centrifuged 3 min at 4000xg. Results: A significantly reduced time duration between reception of sample and available results in informatics lab system was observed with the reduction time of centrifugation allowed by use of Barricor™ tube compared to regular heparin lithium tubes (p < 0.001). A significative decrease in hemolysis rate also occurred in the second period as samples with HI < 10 reached from 52.5% in the first period to 68.5% (p < 0.001) in the second. Conclusion: Low vacuum BarricorTM tubes allowing a higher speed of centrifugation improve lab TAT without impairment of sample quality as a significant reduction of hemolysis was observed, a double advantage which is of particular interest for ED. Keywords: Barricor, Emergency, Hemolysis, Low vacuum, Turnaround time TAT