Sensors (Sep 2016)

Infrared LED Enhanced Spectroscopic CdZnTe Detector Working under High Fluxes of X-rays

  • Jakub Pekárek,
  • Václav Dědič,
  • Jan Franc,
  • Eduard Belas,
  • Martin Rejhon,
  • Pavel Moravec,
  • Jan Touš,
  • Josef Voltr

DOI
https://doi.org/10.3390/s16101591
Journal volume & issue
Vol. 16, no. 10
p. 1591

Abstract

Read online

This paper describes an application of infrared light-induced de-polarization applied on a polarized CdZnTe detector working under high radiation fluxes. We newly demonstrate the influence of a high flux of X-rays and simultaneous 1200-nm LED illumination on the spectroscopic properties of a CdZnTe detector. CdZnTe detectors operating under high radiation fluxes usually suffer from the polarization effect, which occurs due to a screening of the internal electric field by a positive space charge caused by photogenerated holes trapped at a deep level. Polarization results in the degradation of detector charge collection efficiency. We studied the spectroscopic behavior of CdZnTe under various X-ray fluxes ranging between 5 × 10 5 and 8 × 10 6 photons per mm 2 per second. It was observed that polarization occurs at an X-ray flux higher than 3 × 10 6 mm − 2 ·s − 1 . Using simultaneous illumination of the detector by a de-polarizing LED at 1200 nm, it was possible to recover X-ray spectra originally deformed by the polarization effect.

Keywords