Physics and Imaging in Radiation Oncology (Oct 2024)

Reduced-distortion diffusion weighted imaging for head and neck radiotherapy

  • Eric Aliotta,
  • Ramesh Paudyal,
  • Alex Dresner,
  • Amita Shukla-Dave,
  • Nancy Lee,
  • Laura Cerviño,
  • Ricardo Otazo,
  • Victoria Y. Yu

Journal volume & issue
Vol. 32
p. 100653

Abstract

Read online

Background and purpose: Quantitative Diffusion Weighted Imaging (DWI) has potential value in guiding head and neck (HN) cancer radiotherapy. However, clinical translation has been hindered by severe distortions in standard single-shot Echo-Planar-Imaging (ssEPI) and prolonged scan time and low SNR in Turbo-Spin-Echo (ssTSE) sequences. In this study, we evaluate “multi-shot” (ms) msEPI and msTSE acquisitions in the context of HN radiotherapy. Materials and methods: ssEPI, ssTSE, msEPI with 2 and 3 shots (2sEPI, 3sEPI), and msTSE DWI were acquired in a phantom, healthy volunteers (N=10), and patients with HN cancer (N=5) on a 3-Tesla wide-bore MRI in radiotherapy simulation RF coil setup, with matched spatial resolution (2x2x5mm) and b = 0, 200, 800 s/mm2.Geometric distortions measured with deformable vector field (DVF) and contour analysis, apparent diffusion coefficient (ADC) values, and signal-to-noise-ratio efficiency (SNReff) were quantified for all scans. Results: All techniques significantly (P<1x10-3) reduced distortions compared with ssEPI (DVFmean = 3.1 ± 1.3 mm). Distortions were marginally lower for msTSE (DVFmean = 1.5 ± 0.6 mm) than ssTSE (1.8 ± 0.9 mm), but were slightly higher with 2sEPI and 3sEPI (2.6 ± 1.0 mm, 2.2 ± 1.0 mm). SNReff reduced with decreasing distortion with ssEPI=21.9 ± 7.9, 2sEPI=15.1 ± 5.0, 3sEPI=12.1 ± 4.5, ssTSE=6.0 ± 1.6, and msTSE=5.7 ± 1.9 for b = 0 images. Phantom ADC values were consistent across all protocols (errors ≤ 0.03x10-3mm2/s), but in vivo ADC values were ∼ 4 % lower with msEPI and ∼ 12 % lower with ssTSE/msTSE compared with ssEPI. Conclusions: msEPI and TSE acquisitions exhibited improved geometric distortion at the cost of SNReff and scan time. While msTSE exhibited the least distortion, 3sEPI may offer an appealing middle-ground with improved geometric fidelity but superior efficiency and in vivo ADC quantification.