Computer Assisted Surgery (Dec 2024)
Flexible endoscope manipulating robot using quad-roller friction mechanism
Abstract
A robotic system for manipulating a flexible endoscope in surgery can provide enhanced accuracy and usability compared to manual operation. However, previous studies require large-scale, complex hardware systems to implement the rotational and translational motions of the soft endoscope cable. The conventional control of the endoscope by actuating the endoscope handle also leads to undesired slack between the endoscope tip and the handle, which becomes more problematic with long endoscopes such as a colonoscope. This study proposes a compact quad-roller friction mechanism that enables rotational and translational motions triggered not from the endoscope handle but at the endoscope tip. Controlling two pairs of tilted rollers achieves both types of motion within a small space. The proposed system also introduces an unsynchronized motion strategy between the handle and tip parts to minimize the robot’s motion near the patient by employing the slack positively as a control index. Experiments indicate that the proposed system achieves accurate rotational and translational motions, and the unsynchronized control method reduces the total translational motion by up to 88% compared to the previous method.
Keywords