Journal of Biomedical Photonics & Engineering (Apr 2017)

Comparison of a near-infrared reflectance spectroscopy system and skin conductance measurements for in vivo estimation of skin hydration: a clinical study

  • Inga Saknite,
  • Aleksejs Zavorins,
  • Ilona Zablocka,
  • Janis Kisis,
  • Janis Spigulis

DOI
https://doi.org/10.18287/JBPE17.03.010310
Journal volume & issue
Vol. 3, no. 1

Abstract

Read online

Diffuse reflectance spectroscopy system was developed for estimation of skin hydration in the near-infrared spectral range of 900-1700 nm. Experimental setup consisted of a near-infrared spectrometer, Y-type fiber optics probe with 1 detection and 6 illumination fibers, halogen-tungsten light source and a PC. By analyzing diffuse reflectance spectrum, a parameter representing skin hydration by performing baseline correction and calculating the area under the 1450 nm water absorption maximum is proposed. A clinical study was performed acquiring data of skin hydration of 39 patients' forearm skin. Results of the developed system are compared to results obtained by a commercial device based on skin conductance measurements.