Journal of Innovative Optical Health Sciences (May 2022)

Ring artifacts removal in X-ray-induced acoustic computed tomography

  • Prabodh Kumar Pandey,
  • Hari Om Aggrawal,
  • Siqi Wang,
  • Kaitlyn Kim,
  • An Liu,
  • Liangzhong Xiang

DOI
https://doi.org/10.1142/S1793545822500171
Journal volume & issue
Vol. 15, no. 03

Abstract

Read online

X-ray-induced acoustic computed tomography (XACT) is a hybrid imaging modality for detecting X-ray absorption distribution via ultrasound emission. It facilitates imaging from a single projection X-ray illumination, thus reducing the radiation exposure and improving imaging speed. Nonuniform detector response caused by the interference between multichannel data acquisition for ring array transducers and amplifier systems yields ring artifacts in the reconstructed XACT images, which compromises the image quality. We propose model-based algorithms for ring artifacts corrected XACT imaging and demonstrate their efficacy on numerical and experimental measurements. The corrected reconstructions indicate significantly reduced ring artifacts as compared to their conventional counterparts.

Keywords