Journal of Arrhythmia (Jun 2016)

Home monitoring report from a single lead Lumax DX implantable cardioverter defibrillator: New observations in a new system

  • Yuval Konstantino,
  • Alex Kleiman,
  • Guy Amit

DOI
https://doi.org/10.1016/j.joa.2015.12.004
Journal volume & issue
Vol. 32, no. 3
pp. 218 – 219

Abstract

Read online

A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right ventricular channel; these were classified as atrial tachycardia. In the present report, we discussed the unusual pattern of the artifacts that was related to an electromagnetic interference detected by the novel DX system.