Frontiers in Digital Health (Aug 2022)

Accurate detection of heart rate using in-ear photoplethysmography in a clinical setting

  • Tim Adams,
  • Sophie Wagner,
  • Melanie Baldinger,
  • Incinur Zellhuber,
  • Michael Weber,
  • Daniel Nass,
  • Rainer Surges

DOI
https://doi.org/10.3389/fdgth.2022.909519
Journal volume & issue
Vol. 4

Abstract

Read online

BackgroundRecent research has shown that photoplethysmography (PPG) based wearable sensors offer a promising potential for chronic disease monitoring. The aim of the present study was to assess the performance of an in-ear wearable PPG sensor in acquiring valid and reliable heart rate measurements in a clinical setting, with epileptic patients.MethodsPatients undergoing video-electroencephalography (EEG) monitoring with concomitant one-lead electrocardiographic (ECG) recordings were equipped with an in-ear sensor developed by cosinuss°.ResultsIn total, 2,048 h of recording from 97 patients with simultaneous ECG and in-ear heart rate data were included in the analysis. The comparison of the quality-filtered in-ear heart rate data with the reference ECG resulted in a bias of 0.78 bpm with a standard deviation of ±2.54 bpm; Pearson’s Correlation Coefficient PCC = 0.83; Intraclass Correlation Coefficient ICC = 0.81 and mean absolute percentage error MAPE = 2.57.ConclusionThese data confirm that the in-ear wearable PPG sensor provides accurate heart rate measurements in comparison with ECG under realistic clinical conditions, especially with a signal quality indicator. Further research is required to investigate whether this technology is helpful in identifying seizure-related cardiovascular changes.

Keywords