Diagnostics (Sep 2023)

Comparison of the 11-Day Adhesive ECG Patch Monitor and 24-h Holter Tests to Assess the Response to Antiarrhythmic Drug Therapy in Paroxysmal Atrial Fibrillation

  • Soohyun Kim,
  • Young Choi,
  • Kichang Lee,
  • Sung-Hwan Kim,
  • Hwajung Kim,
  • Sanghoon Shin,
  • Soyoon Park,
  • Yong-Seog Oh

DOI
https://doi.org/10.3390/diagnostics13193078
Journal volume & issue
Vol. 13, no. 19
p. 3078

Abstract

Read online

Accurate assessment of the response to the antiarrhythmic drug (AAD) in atrial fibrillation (AF) is crucial to achieve adequate rhythm control. We evaluated the effectiveness of extended cardiac monitoring using an adhesive ECG patch in the detection of drug-refractory paroxysmal AF. Patients diagnosed with paroxysmal AF and receiving AAD therapy were enrolled. The subjects simultaneously underwent 11-day adhesive ECG patch monitoring and a 24-h Holter test. The primary study outcome was a detection rate of drug-refractory AF or atrial tachycardia (AT) lasting ≥30 s. A total of 59 patients were enrolled and completed the study examinations. AF or AT was detected in 28 (47.5%) patients by an 11-day ECG patch monitor and in 8 (13.6%) patients by a 24-h Holter test (p < 0.001). The 11-day ECG patch monitor identified an additional 20 patients (33.8%) with drug-refractory AF not detected by the 24-h Holter, and as a result, the treatment plan was changed in 11 patients (10 catheter ablations, one medication change). In conclusion, extended cardiac rhythm monitoring using an adhesive ECG patch in patients with paroxysmal AF under AAD therapy led to over a threefold higher detection of drug-refractory AF episodes, compared to the 24-h Holter test.

Keywords