Journal of Innovative Optical Health Sciences (Nov 2020)

Stick-slip nonuniform rotation distortion correction in distal scanning optical coherence tomography catheters

  • Jessica Mavadia-Shukla,
  • Jianlin Zhang,
  • Kaiyan Li,
  • Xingde Li

DOI
https://doi.org/10.1142/S1793545820500303
Journal volume & issue
Vol. 13, no. 6
pp. 2050030-1 – 2050030-8

Abstract

Read online

We present a robust and fiducial-marker-free algorithm that can identify and correct stick-slip distortion caused by nonuniform rotation (or beam scanning) in distally scanned catheters for endoscopic optical coherence tomography (OCT) images. This algorithm employs spatial frequency analysis to select and remove distortions. We demonstrate the feasibility of this algorithm on images acquired from ex vivo rat colon with a distally scanned DC motor-based endoscope. The proposed algorithm can be applied to general endoscopic OCT images for correcting nonuniform rotation distortion.

Keywords