Journal of the Serbian Chemical Society (Jan 2005)

Determination of boron in the waters of Troia by inductively coupled plasma-atomic emission spectrometry (ICP-AES)

  • Ozcan Hasan,
  • Yilmaz Selahattin

DOI
https://doi.org/10.2298/jsc0510219o
Journal volume & issue
Vol. 70, no. 10
pp. 1219 – 1227

Abstract

Read online

The inductively coupled plasma-atomic emission spectrometry method (ICP-EAS) is proposed for the determination of boron in Troia water sources. In this study, boron samples were taken from 25 different locations during November 2002 and May 2003. The 249.773 nm boron emmision line was not affected by the sample matrix. The boron content was determined without further pretreatment of samples proposed ICP-AES method. An analysis of a given sample was completed in about 15 min. The determined boron concentrations in surface and ground water sources were between 0.00444–1.11200 mg l-1. In order to compare the results obtained by the proposed method, selected water samples were also analysed titrimetically. Pearson correlation coefficient analysis showed that there is a close relationship between the ICP-AES and the titrimetric method and the determination coefficient (R2) was calculated to be 91.34 %.