BMC Medical Research Methodology (Jun 2007)

Comparison of two different physical activity monitors

  • Baer David J,
  • Moshfegh Alanna J,
  • Kramer Matthew,
  • Paul David R,
  • Rumpler William V

DOI
https://doi.org/10.1186/1471-2288-7-26
Journal volume & issue
Vol. 7, no. 1
p. 26

Abstract

Read online

Abstract Background Understanding the relationships between physical activity (PA) and disease has become a major area of research interest. Activity monitors, devices that quantify free-living PA for prolonged periods of time (days or weeks), are increasingly being used to estimate PA. A range of different activity monitors brands are available for investigators to use, but little is known about how they respond to different levels of PA in the field, nor if data conversion between brands is possible. Methods 56 women and men were fitted with two different activity monitors, the Actigraph™ (Actigraph LLC; AGR) and the Actical™ (Mini-Mitter Co.; MM) for 15 days. Both activity monitors were fixed to an elasticized belt worn over the hip, with the anterior and posterior position of the activity monitors randomized. Differences between activity monitors and the validity of brand inter-conversion were measured by t-tests, Pearson correlations, Bland-Altman plots, and coefficients of variation (CV). Results The AGR detected a significantly greater amount of daily PA (216.2 ± 106.2 vs. 188.0 ± 101.1 counts/min, P Conclusion Although activity monitors predict PA on the same scale (counts/min), the results between these two brands are not directly comparable. However, the data are comparable if a conversion equation is applied, with better results for log-transformed data.