The Egyptian Journal of Radiology and Nuclear Medicine (Apr 2024)

Low-dose versus standard-dose normal temporal bone CT in children: a comparison study

  • R. Rashma,
  • Jyoti Kumar,
  • Anju Garg,
  • Radhika Batra,
  • Ravi Meher,
  • Ankita Phulia

DOI
https://doi.org/10.1186/s43055-024-01254-7
Journal volume & issue
Vol. 55, no. 1
pp. 1 – 15

Abstract

Read online

Abstract Objective To compare the image quality of normal anatomical structures and radiation dose on low-dose (LDCT) and standard-dose (SDCT) temporal bone CT in children. Methods The study included 45 LDCT (80 kV and 130 mAs) and 45 SDCT (120 kV and 170 mAs) scans in children, 1–15 years of age. LDCT and SDCT scans were analyzed on H60s and H70h reconstruction kernels, respectively. Two readers assessed the image quality for 25 anatomical structures, using a 5-point scale. A score of 3 and above was considered “sufficient” and 2 and below was considered “insufficient” image quality. Image noise, contrast, age and size-specific effective doses were calculated. Results Despite an increase in image noise on LDCT, image quality remained sufficient for most structures owing to increased image contrast. The median effective dose on LDCT, calculated with age-specific conversion factor, decreased by 72.9% and that calculated with size-specific conversion factor decreased by 81.8% compared to the dose on SDCT. Conclusion LDCT provides comparable image quality for evaluation of temporal bone with significant reduction in radiation dose in children.