Physics and Imaging in Radiation Oncology (Oct 2020)

Dose-response of deformable radiochromic dosimeters for spot scanning proton therapy

  • Simon V. Jensen,
  • Lia B. Valdetaro,
  • Per R. Poulsen,
  • Peter Balling,
  • Jørgen B.B. Petersen,
  • Ludvig P. Muren

Journal volume & issue
Vol. 16
pp. 134 – 137

Abstract

Read online

Intrafractional motion and deformation influence proton therapy delivery for tumours in the thorax, abdomen and pelvis. This study aimed to test the dose–response of a compressively strained three-dimensional silicone-based radiochromic dosimeter during proton beam delivery. The dosimeter was read-out in its relaxed state using optical computed tomography and calibrated for the linear energy transfer, based on Monte Carlo simulations. A three-dimensional gamma analysis showed a 99.3% pass rate for 3%/3 mm and 93.9% for 2%/2 mm, for five superimposed measurements using deformation-including Monte Carlo dose calculations as reference. We conclude that the dosimeter’s dose–response is unaffected by deformations.