Toxics (Sep 2023)
Assessment of Radiation Exposure in a Nuclear Medicine Department during <sup>99m</sup>Tc-MDP Bone Scintigraphy
Abstract
This study measured 99mTc-MDP bone scintigraphy radiation risks, as low-dose radiation exposure is a growing concern. Dosimeter measurements were taken at four positions (left lateral, right lateral, anterior, and posterior) around the patients at 30, 60, 100, and 200 cm at 0, 1.5, and 3 h. The highest dose rates were recorded from 51% of the patients, who emitted ≥ 25 µSv/h up to 49.00 µSv/h at the posterior location at a distance of 30 cm. Additionally, at the anterior location at a distance of 30 cm, 42% of patients emitted ≥ 25 µSv/h up to 38.00 µSv/h. Furthermore, at 1.5 h after the tracer injection, 7% of the dose rates exceeded 25 µSv/h. There was a significant reduction in mean dose rates for all positions as distance and time increased (p-value 99mTc-MDP were below occupational limits. Thus, medical staff received less radiation than the recommended 25 μSv/h. On discharge and release to public areas, the patients’ mean dose rates were as follows: 1.13 µSv/h for the left lateral position, 1.04 µSv/h for the right lateral, 1.39 µSv/h for the anterior, and 1.46 µSv/h for the posterior. This confirms that if an individual was continuously present in an unrestricted area, the dose from external sources would not exceed 20 µSv/h. Furthermore, the patients’ radiation doses were below the public exposure limit on discharge.