Proceedings (Nov 2020)

Droplet Reactors with Bioluminescent Enzymes for Real-Time Water Pollution Monitoring

  • Anton S. Yakimov,
  • Ivan A. Denisov,
  • Anton S. Bukatin,
  • Kirill A. Lukyanenko,
  • Kirill I. Belousov,
  • Igor V. Kukhtevich,
  • Elena N. Esimbekova,
  • Anatoly A. Evstrapov,
  • Peter I. Belobrov

DOI
https://doi.org/10.3390/IECB2020-07046
Journal volume & issue
Vol. 60, no. 1
p. 54

Abstract

Read online

Early detection of pollutants in wastewater, water coming out of treatment facilities, drinking water, and water for agricultural needs is a challenging problem. Effective water quality monitoring requires development of new methods for express detection of pollutants. Enzymes from bioluminescent bacteria can be used for the development of new express enzyme-based bioassay systems. This work demonstrates, for the first time, a microfluidic chip to generate emulsion droplets containing two enzymes of the bacterial bioluminescent system (luciferase and NAD(P)H:FMN-oxidoreductase) with reaction substrates. The developed chip generated “water-in-oil” emulsion droplets with a volume of 0.1 μL and a frequency of up to 12 droplets per second. A portable photomultiplier tube (PMT) was used to measure the bioluminescent signal in each individual droplet; the signal-to-noise ratio was 3000/1. The intensity of luminescence in droplets depended on the concentration of copper ions. The limit of detection (LOD) for copper sulfate was 1 mg/L. We showed that bioluminescent enzymatic reactions can be carried out in droplet reactors that can be applied for online monitoring of water quality. Thus, the suggested method of biological measurements has a good perspective for biosensing in general.