Influence of reverse total shoulder arthroplasty baseplate design on torque and compression relationship

  • Miguel A. Diaz, MS,
  • Jason E. Hsu, MD,
  • Eric T. Ricchetti, MD,
  • Grant E. Garrigues, MD,
  • Sergio Gutierrez, PhD,
  • Mark A. Frankle, MD

Journal volume & issue
Vol. 4, no. 2
pp. 388 – 396

Abstract

Read online

Background: A linear relationship between baseplate insertion torque and compression force in reverse shoulder arthroplasty (RSA) baseplates with central screw design has been recently established. In this study, we evaluated 3 different baseplate designs and their influence on the torque-compression relationship. Methods: Three different RSA baseplate designs were evaluated through biomechanical testing using a glenoid vault, bone surrogate model. A digital torque gauge was used to measure insertion torque applied to the baseplate, whereas compression data were collected continuously from a load cell. Additionally, 2 predictive models were developed to predict the compression forces of each baseplate design at varying levels of torque. Results: A linear relationship was found between baseplate compression and insertion torque for all 3 baseplate designs. Both the monoblock and 2-piece locking designs achieved the goal torque of 6.8 Nm, whereas the 2-piece nonlocking design did not due to material strip-out. No significant difference in maximum compression was found between the monoblock and 2-piece locking designs. However, the 2-piece nonlocking design achieved significantly higher compression. Both predictive models were shown to adequately predict compressive forces at different torque inputs for the monoblock and 2-piece locking designs but not the 2-piece nonlocking design. Conclusion: The torque-compression relationship of a central screw baseplate is significantly affected by baseplate design. A 2-piece nonlocking baseplate reaches higher compression levels and risks material strip-out at lower insertional torques compared with a monoblock and 2-piece locking design. This has implications both on component design and on surgeon tactile feedback during surgery.

Keywords