BMC Musculoskeletal Disorders (Jan 2024)

Ultrasound had high accuracy in measuring hip joint capsule thickness

  • Guanying Gao,
  • Huaan Fang,
  • Kaicheng Zhou,
  • Zizhi Mo,
  • Jiayang Liu,
  • Lingyu Meng,
  • Jianquan Wang,
  • Yan Xu

DOI
https://doi.org/10.1186/s12891-024-07228-0
Journal volume & issue
Vol. 25, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Background The hip joint capsule is an essential component of hip joint function and stability, and its thickness is closely associated with certain medical conditions, surgical outcomes, and rehabilitation treatments. Currently, in clinical practice, hip joint capsule thickness is predominantly measured using magnetic resonance imaging (MRI), with limited utilization of ultrasound examinations for this purpose. Methods We retrospectively evaluated patients who visited our Sports Medicine Department between February 2017 and March 2023 and underwent both hip joint MRI and ultrasound imaging on the same side. All patients had undergone preoperative hip joint MRI and ultrasound examinations, with the time gap between the two examinations not exceeding three months. Measurements of hip joint capsule thickness were taken on both MRI and ultrasound images for the same patients to analyze their consistency. Additionally, we measured the alpha angle, lateral center-edge angle, acetabular anteversion angle, and femoral anteversion angle of the patients’ hip joints and analyzed their correlation with hip joint capsule thickness measure by ultrasound. Results A total of 307 patients were included in this study, with hip joint capsule thickness measured by MRI and ultrasound being 5.0 ± 1.2 mm and 5.0 ± 1.5 mm, respectively. The Bland-Altman analysis demonstrates good agreement or consistency. The paired t-test resulted in a p-value of 0.708, indicating no significant statistical difference between the two methods. The correlation analysis between acetabular anteversion angle and ultrasound-measured capsule thickness yielded a p-value of 0.043, indicating acetabular anteversion angle and capsular thickness may have negative correlation. Conclusions The measurements of joint capsule thickness obtained through ultrasound and MRI showed good consistency, suggesting that ultrasound can be used in clinical practice as a replacement for MRI in measuring hip joint capsule thickness. There was a significant correlation between acetabular anteversion angle and hip joint capsule thickness, indicating potential for further research in this area.

Keywords