IDH1 and IDH2 somatic mutations have been identified in solid tumors and blood malignancies. The development of inhibitors of mutant IDH1 and IDH2 in the past few years has prompted the development of a fast and sensitive assay to detect IDH1R132, IDH2R140 and IDH2R172 mutations to identify patients eligible for these targeted therapies. This study aimed to compare two new multiplexed PCR assays – an automated quantitative PCR (qPCR) on the PGX platform and a droplet digital PCR (ddPCR) with next‐generation sequencing (NGS) for IDH1/2 mutation detection. These assays were evaluated on 102 DNA extracted from patient peripheral blood, bone marrow and formalin‐fixed paraffin‐embedded tissue samples with mutation allelic frequency ranging from 0.6% to 45.6%. The ddPCR assay had better analytical performances than the PGX assay with 100% specificity, 100% sensitivity and a detection limit down to 0.5% on IDH1R132, IDH2R140 and IDH2R172 codons, and a high correlation with NGS results. Therefore, the new highly multiplexed ddPCR is a fast and cost‐effective assay that meets most clinical needs to identify and follow cancer patients in the era of anti‐IDH1/2‐targeted therapies.