BMC Musculoskeletal Disorders (Jul 2017)

The effect of different screw-rod design on the anti-rotational torque: a biomechanical comparison of three conventional screw-rod constructs

  • Zifang Huang,
  • Chongwen Wang,
  • Hengwei Fan,
  • Wenyuan Sui,
  • Xueshi Li,
  • Qifei Wang,
  • Junlin Yang

DOI
https://doi.org/10.1186/s12891-017-1683-1
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 5

Abstract

Read online

Abstract Background Screw-rod constructs have been widely used to correct spinal deformities, but the effects of different screw-rod systems on anti-rotational torque have not been determined. This study aimed to analyze the biomechanical effect of different rod-screw constructs on anti-rotational torque. Methods Three conventional spinal screw-rod systems (Legacy, RF-F-10 and USSII) were used to test the anti-rotational torque in the material test machine. ANOVA was performed to evaluate the anti-rotational capacity of different pedicle screws-rod constructs. Results The anti-rotational torque of Legacy group, RF-F-10 group and USSII group were 12.3 ± 1.9 Nm, 6.8 ± 0.4 Nm, and 3.9 ± 0.8 Nm, with a P value lower than 0.05. This results indicated that the Legacy screws-rod construct could provide a highest anti-rotation capacity, which is 68% and 210% greater than RF-F-10 screw-rod construct and USSII screw-rod respectively. Conclusions The anti-rotational torque may be mainly affected by screw cap and groove design. Our result showed the anti-rotational torque are: Legacy system > RF-F-10 system > USSII system, suggesting that appropriate rod-screw constructs selection in surgery may be vital for anti-rotational torque improvement and preventing derotation correction loss.