Abstract

Read online

Introduction: Despite considerable benefits associated with current drug-eluting stents, continued attention to the safety, efficacy, and deliverability of available drug-eluting stent has led to the development of newer stent. Methods: This study was a single-centre, prospective, non-randomized, first-in-man study which included clinical follow-up data was collected at 1, 8 and 12 months after the procedure. The study included 105 patients with de novo native coronary artery lesions including multi-vessel disease treated with Supralimus-Core® stent. Repeat angiography was performed 8 months post-stent implantation. Results: At quantitative coronary angiography 8-month luminal late loss was 0.39 ± 0.33 mm in-stent and 0.33 ± 0.35 mm in-segment. The incidence of any major adverse cardiac event at 30 days, 8 months and 12 months was 1 (1%), 6 (6%) and 7 (7%) respectively. Conclusion: This study demonstrates that the Supralimus-Core® SES is a safe and effective treatment for patients with obstructive coronary artery disease. ClinicalTrials.gov ID: NCT00811616.