Sensors (Mar 2018)

Eight-Channel AC Magnetosusceptometer of Magnetic Nanoparticles for High-Throughput and Ultra-High-Sensitivity Immunoassay

  • Jen-Jie Chieh,
  • Wen-Chun Wei,
  • Shu-Hsien Liao,
  • Hsin-Hsein Chen,
  • Yen-Fu Lee,
  • Feng-Chun Lin,
  • Ming-Hsien Chiang,
  • Ming-Jang Chiu,
  • Herng-Er Horng,
  • Shieh-Yueh Yang

DOI
https://doi.org/10.3390/s18041043
Journal volume & issue
Vol. 18, no. 4
p. 1043

Abstract

Read online

An alternating-current magnetosusceptometer of antibody-functionalized magnetic nanoparticles (MNPs) was developed for immunomagnetic reduction (IMR). A high-sensitivity, high-critical-temperature superconducting quantum interference device was used in the magnetosusceptometer. Minute levels of biomarkers of early-stage neurodegeneration diseases were detectable in serum, but measuring each biomarker required approximately 4 h. Hence, an eight-channel platform was developed in this study to fit minimal screening requirements for Alzheimer’s disease. Two consistent results were measured for three biomarkers, namely Aβ40, Aβ42, and tau protein, per human specimen. This paper presents the instrument configuration as well as critical characteristics, such as the low noise level variations among channels, a high signal-to-noise ratio, and the coefficient of variation for the biomarkers’ IMR values. The instrument’s ultrahigh sensitivity levels for the three biomarkers and the substantially shorter total measurement time in comparison with the previous single- and four-channels platforms were also demonstrated in this study. Thus, the eight-channel instrument may serve as a powerful tool for clinical high-throughput screening of Alzheimer’s disease.

Keywords