EPJ Web of Conferences (Jan 2022)

In-vitro measurement of aerosol generated by Respimat SMI penetrating through upper airways

  • Misik Ondrej,
  • Lizal Frantisek,
  • Malý Milan,
  • Karas Jakub,
  • Jedelsky Jan,
  • Jicha Miroslav

DOI
https://doi.org/10.1051/epjconf/202226901038
Journal volume & issue
Vol. 269
p. 01038

Abstract

Read online

Respimat soft mist inhaler (SMI) is an inhaler producing a soft mist with low velocity. The liquid drug solution is emitted by the mechanical energy of compressed spring, so the device is propellant-free. This paper describes the measurement of particle fraction penetrating through the model of the upper respiratory tract (mouth cavity, pharynx, larynx trachea). The solution inside the cartridge was marked by fluorescein in order to evaluate the regional particle deposition in the model. After the exposition, segments of the model, as well as the filter, were washed out and the deposited fluorescein was extracted into water. The analysis showed that the fraction of particles penetrating through the upper respiratory tract was 64.7±6.7 %. However, the real inhalation pattern was not simulated and measurement was provided with continual “inhalation” without “exhalation”.