Journal of Arrhythmia (Aug 2023)

Compatibility of a novel temperature‐controlled, irrigated radiofrequency catheter with ultra‐high‐density mapping

  • Daniel Rodriguez Muñoz,
  • Álvaro Marco del Castillo,
  • Javier Ramos Jimenez,
  • Luis Borrego Bernabe,
  • Alba Madrid Montoya,
  • Adrián Lorenzo Balboa,
  • Fernando Arribas Ynsaurriaga,
  • Rafael Salguero‐Bodes

DOI
https://doi.org/10.1002/joa3.12896
Journal volume & issue
Vol. 39, no. 4
pp. 676 – 680

Abstract

Read online

Abstract Background Compatibility of DiamondTemp (DT) radiofrequency (RF) catheter with the Rhythmia mapping system has not been manufacturer‐reported nor its tracking accuracy reported. Methods Consecutive patients undergoing macroreentrant atrial tachycardia ablation guided by Rhythmia and ablated using DT were prospectively enrolled. Following catheter configuration, ablation lines were performed and remapped to measure the RF tag to effective‐ablation‐line‐center (RFT‐ALC) distance. Results Among 20 consecutive patients (54 maps), 40 ablation lines were evaluated. Overall, the RFT‐ALC distance was 3.88 ± 2.95 mm, and the operator assessment of accuracy was high. No complications occurred. Conclusion The use of DT catheter guided by the Rhythmia mapping system is feasible and accurate.

Keywords