Journal of Pharmaceutical Analysis (Feb 2014)
LCâUV/MS quality analytics of paediatric artemether formulations
Abstract
A highly selective and stability-indicating HPLC-method, combined with appropriate sample preparation steps, is developed for β-artemether assay and profiling of related impurities, including possible degradants, in a complex powder for oral suspension. Following HPLC conditions allowed the required selectivity: a Prevail organic acid (OA) column (250 mmÃ4.6 mm, 5 μm), flow rate set at 1.5 mL/min combined with a linear gradient (where A=25 mM phosphate buffer (pH 2.5), and B=acetonitrile) from 30% to 75% B in a runtime of 60 min. Quantitative UV-detection was performed at 210 nm. Acetonitrile was applied as extraction solvent for sample preparation. Using acetonitrileâwater mixtures as extraction solvent, a compartmental behaviour by a non-solving excipient-bound fraction and an artemether-solubilising free fraction of solvent was demonstrated, making a mobile phase based extraction not a good choice. Method validation showed that the developed HPLC-method is considered to be suitable for its intended regulatory stability-quality characterisation of β-artemether paediatric formulations. Furthermore, LCâMS on references as well as on stability samples was performed allowing identity confirmation of the β-artemether related impurities. MS-fragmentation scheme of β-artemether and its related substances is proposed, explaining the m/z values of the in-source fragments obtained. Keywords: Paediatric formulations, Sample preparation, Polar embedded organic acid column, Artemisinin trioxane derivatives, Related impurities and degradation compounds, MS-fragmentation